Infrared hyperspectral imaging stokes polarimeter
by Jones, Julia Craven, Ph.D., THE UNIVERSITY OF ARIZONA, 2011, 127 pages; 3450734

Abstract:

This work presents the design, development, and testing of a field portable imaging spectropolarimeter that operates over the short-wavelength and middle-wavelength portion of the infrared spectrum. The sensor includes a pair of sapphire Wollaston prisms and several high order retarders to produce the first infrared implementation of an imaging Fourier transform spectropolarimeter, providing for the measurement of the complete spectropolarimetric datacube over the passband. The Wollaston prisms serve as a birefringent interferometer with reduced sensitivity to vibration when compared to an unequal path interferometer, such as a Michelson. Polarimetric data are acquired through the use of channeled spectropolarimetry to modulate the spectrum with the Stokes parameter information. The collected interferogram is Fourier filtered and reconstructed to recover the spatially and spectrally varying Stokes vector data across the image.

The intent of this dissertation is to provide the reader with a detailed understanding of the steps involved in the development of this infrared hyperspectral imaging polarimeter (IHIP) instrument. First, Chapter 1 provides an overview of the fundamental concepts relevant to this research. These include imaging spectrometers, polarimeters, and spectropolarimeters. A detailed discussion of channeled spectropolarimetry, including a historical study of previous implementations, is also presented. Next a few of the design alternatives that are possible for this work are outlined and discussed in Chapter 2. The configuration that was selected for the IHIP is then presented in detail, including the optical layout, design, and operation. Chapter 3 then presents an artifact reduction technique (ART) that was developed to improve the IHIP’s spectropolarimetric reconstructions by reducing errors associated with non-band-limited spectral features. ART is experimentally verified in the infrared using a commercial Fourier transform spectrometer in combination with Yttrium Vanadate as well as Cadmium Sulfide retarders.

The remainder of this dissertation then details the testing and analysis of the IHIP instrument. Implementation of ART with the IHIP as well as the employed calibration techniques are described in Chapter 4. Complete calibration of the IHIP includes three distinct processes to provide radiometric, spectral, and polarimetric calibration. With the instrument assembled and calibrated, results and error analyses are presented in Chapter 5. Spectropolarimetric results are obtained in the laboratory as well as outdoors to test the IHIP’s real world functionality. The performance of the instrument is also assessed, including experimental measurement of signal-to-noise ratio (SNR), and an analysis of the potential sources of systematic error (such as retarder misalignment and finite polarizer extinction ratio). Chapter 6 presents the design and experimental results for a variable Wollaston prism that can be added to the IHIP to vary the fringe contrast across the field of view. Finally, Chapter 7 includes brief closing remarks summarizing this work and a few observations which may be useful for future infrared imaging Fourier transform channeled spectropolarimeter instruments.

 
AdviserEustace L. Dereniak
SchoolTHE UNIVERSITY OF ARIZONA
SourceDAI/B 72-07, p. , May 2011
Source TypeDissertation
SubjectsOptics
Publication Number3450734
Adobe PDF Access the complete dissertation:
 

» Find an electronic copy at your library.
  Use the link below to access a full citation record of this graduate work:
  http://gateway.proquest.com/openurl%3furl_ver=Z39.88-2004%26res_dat=xri:pqdiss%26rft_val_fmt=info:ofi/fmt:kev:mtx:dissertation%26rft_dat=xri:pqdiss:3450734
  If your library subscribes to the ProQuest Dissertations & Theses (PQDT) database, you may be entitled to a free electronic version of this graduate work. If not, you will have the option to purchase one, and access a 24 page preview for free (if available).

About ProQuest Dissertations & Theses
With over 2.3 million records, the ProQuest Dissertations & Theses (PQDT) database is the most comprehensive collection of dissertations and theses in the world. It is the database of record for graduate research.

The database includes citations of graduate works ranging from the first U.S. dissertation, accepted in 1861, to those accepted as recently as last semester. Of the 2.3 million graduate works included in the database, ProQuest offers more than 1.9 million in full text formats. Of those, over 860,000 are available in PDF format. More than 60,000 dissertations and theses are added to the database each year.

If you have questions, please feel free to visit the ProQuest Web site - http://www.proquest.com - or call ProQuest Hotline Customer Support at 1-800-521-3042.