Abstract:

In this work, three methods of measuring the polarization state of light in the thermal infrared (3-12 μm) are modeled, simulated, calibrated and experimentally verified in the laboratory. The first utilizes the method of channeled spectropolarimetry (CP) to encode the Stokes polarization parameters onto the optical power spectrum. This channeled spectral technique is implemented with the use of two Yttrium Vanadate (YVO4) crystal retarders. A basic mathematical model for the system is presented, showing that all the Stokes parameters are directly present in the interferogram. Theoretical results are compared with real data from the system, an improved model is provided to simulate the effects of absorption within the crystal, and a modified calibration technique is introduced to account for this absorption. Lastly, effects due to interferometer instabilities on the reconstructions, including non-uniform sampling and interferogram translations, are investigated and techniques are employed to mitigate them.

Second is the method of prismatic imaging polarimetry (PIP), which can be envisioned as the monochromatic application of channeled spectropolarimetry. Unlike CP, PIP encodes the 2-dimensional Stokes parameters in a scene onto spatial carrier frequencies. However, the calibration techniques derived in the infrared for CP are extremely similar to that of the PIP. Consequently, the PIP technique is implemented with a set of four YVO4 crystal prisms. A mathematical model for the polarimeter is presented in which diattenuation due to Fresnel effects and dichroism in the crystal are included. An improved polarimetric calibration technique is introduced to remove the diattenuation effects, along with the relative radiometric calibration required for the BPIP operating with a thermal background and large detector offsets. Data demonstrating emission polarization are presented from various blackbodies, which are compared to data from our Fourier transform infrared spectropolarimeter. Additionally, limitations in the PIP technique with regards to the spectral bandwidth and F/# of the imaging system are analyzed. A model able to predict the carrier frequency's fringe visibility is produced and experimentally verified, further reinforcing the PIP's limitations.

The last technique is significantly different from CP or PIP and involves the simulation and calibration of a thermal infrared division of amplitude imaging Stokes polarimeter. For the first time, application of microbolometer focal plane array (FPA) technology to polarimetry is demonstrated. The sensor utilizes a wire-grid beamsplitter with imaging systems positioned at each output to analyze two orthogonal linear polarization states simultaneously. Combined with a form birefringent wave plate, the system is capable of snapshot imaging polarimetry in any one Stokes vector (S1, S2 or S3). Radiometric and polarimetric calibration procedures for the instrument are provided and the reduction matrices from the calibration are compared to rigorous coupled wave analysis (RCWA) and raytracing simulations. The design and optimization of the sensor's wire-grid beam splitter and wave plate are presented, along with their corresponding prescriptions. Polarimetric calibration error due to the spectrally broadband nature of the instrument is also overviewed. Image registration techniques for the sensor are discussed and data from the instrument are presented, demonstrating a microbolometer's ability to measure the small intensity variations corresponding to polarized emission in natural environments.

 
AdviserEustace L. Dereniak
SchoolTHE UNIVERSITY OF ARIZONA
SourceDAI/B 70-04, p. , Jul 2009
Source TypeDissertation
SubjectsOptics
Publication Number3354932
Adobe PDF Access the complete dissertation:
 

» Find an electronic copy at your library.
  Use the link below to access a full citation record of this graduate work:
  http://gateway.proquest.com/openurl%3furl_ver=Z39.88-2004%26res_dat=xri:pqdiss%26rft_val_fmt=info:ofi/fmt:kev:mtx:dissertation%26rft_dat=xri:pqdiss:3354932
  If your library subscribes to the ProQuest Dissertations & Theses (PQDT) database, you may be entitled to a free electronic version of this graduate work. If not, you will have the option to purchase one, and access a 24 page preview for free (if available).

About ProQuest Dissertations & Theses
With over 2.3 million records, the ProQuest Dissertations & Theses (PQDT) database is the most comprehensive collection of dissertations and theses in the world. It is the database of record for graduate research.

The database includes citations of graduate works ranging from the first U.S. dissertation, accepted in 1861, to those accepted as recently as last semester. Of the 2.3 million graduate works included in the database, ProQuest offers more than 1.9 million in full text formats. Of those, over 860,000 are available in PDF format. More than 60,000 dissertations and theses are added to the database each year.

If you have questions, please feel free to visit the ProQuest Web site - http://www.proquest.com - or call ProQuest Hotline Customer Support at 1-800-521-3042.